مقال من تأليف: Azgui, F. ; Mears, C. A. ; Labov, S. E. ; Frank, M. ; Sadoulet, B. ; Lindeman, M. ; Golwala, S. ; Hiller, L. J. ; Netel, H. ;

ملخص: Measurements of a Normal metal-Insulating-Superconducting (NIS) tunnel junction sensor Ag/AlOx/Al coupled to a thick Pb absorber are presented. All films were thermally deposited on a Si3N4 membrane using a shadow-mask technique. Irradiation experiments with a 55Fe source were performed using a high bandwidth SQUID with one junction only. We observe a strong position dependence of the pulse shape and amplitude. The different classes of events are probably related to different elements of the sensor. At a bath temperature of 80 mK we have obtained for the main Mn K[alpha] peak an energy resolution of 100 eV FWHM. We have preliminary evidence for absorption in the lead and efficient trapping of the resulting quasiparticles on the NIS normal electrode.